16・32・64・128 ChannelDigital EEG for Research
Bioelectric potential measuring device
The Third Generation
Active EEG System

| Specifications | ||||
|---|---|---|---|---|
| EEG channel | 16 ch, 32 ch, 64 ch, 128 ch | |||
| EMG channel | 16 ch, 32 ch, 64 ch, 128 ch *The electrode tip has been changed to a flat electrode. | |||
| Optional channel | Respiration, GSR, body temperature, pulse wave, pressure transducer, optical sensor (each sensor is automatically recognized) | |||
| Bioelectric potential measurement | 8 terminals (bipolar lead max. 4 ch/reference electrode lead max. 8 ch) | |||
| Sampling rate | 16,384Hz (independent of channel number) adjastable downwards via software | |||
| A/D resolution | 24-bit SAR converter per channel | |||
| Frequency bandwidth (-3 dB) | DC to 5.4 kHz | |||
| Quantization Resolution | LSB = 31.25nV | |||
| Input impedance | 600 MΩ@ 50 Hz | |||
| Input range | +200 mV to -200 mV | |||
| CMR | > 160 dB @ 50 Hz | |||
| Trigger input | 16 ch / TTL level | |||
| External analog signal input | Max. 6 ch (analog/audio signal input code required) | |||
| Power supply | 3,500 mAh, 3.7 V Li-ion battery Type 18650 | |||